Guidance And Manufacturer's Declaration - Electromagnetic Immunity - Atmos C 051 Thorax Operating Instructions Manual

Table of Contents

Advertisement

13.2
Guidance and manufacturer's declaration - electromagnetic
immunity
The ATMOS® C 051 Thorax is intended for use in the electromagnetic environment specified
below. The customer or user of the ATMOS® C 051 Thorax should ensure that it is used in such
an environment.
IEC 60601- Test
Immunity Test
Level
Electrostatic
± 6 kV Contact
discharge (ESD)
± 8 kV Air
according to IEC
61000-4-2
Fast electrical
± 2 kV Mains
transient/burst
± 1 kV I/Os
IEC 61000-4-4
Surges
1 kV
IEC 61000-4-5
Differential
2 kV
Common
Magnetic field at
3 A/m
power frequency
50/60 Hz acc. to
IEC 61000-4-8
Voltage Dips,
< 5 % UT
dropout and
(> 95 % Dip of the
fluctuations
UT) for 0.5 cycles
in the supply
voltage acc. to
40 % UT
IEC 61000-4-11
(60% Dip of the UT)
for 5 cycles
70% UT
(30 % Dip of the
UT) for 25 cycles
< 5 % UT
(>95 % Dip of the
UT) for 5 s
NOTE
UT is the alternating maims voltage prior to application of the test levels.
Compliance Level
± 6 kV Contact
± 8 kV Air
± 2 kV inapplicable
for power cables
± 1 kV I/Os
1 kV
Differential
2 kV
Common
applicable
3 A/m
< 5 % UT
(> 95 % Dip of the
UT) for 0.5 cycles
40 % UT
(60% Dip of the UT)
for 5 cycles
70% UT
(30 % Dip of the UT)
for 25 cycles
< 5 % UT
(>95 % Dip of the UT)
for 5 s
Electromagnetic
Environment - Guidance
Floors should be wood,
concrete, or ceramics tile.
If floors are synthetic, the
relative humidity should be at
least 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency magnetic
fields should be that of a
typical commercial or hospital
environment.
Mains power quality should be
that of a typical commercial
or hospital environment. If
the user of the ATMOS® C 051
Thorax demands continued
function even in case of
interruptions of the energy
supply, it is recommended
to supply the ATMOS® C 051
Thorax from an uninterruptible
current supply or a battery.
53