Secretion Canister - Atmos C 051 Thorax Operating Instructions Manual

Table of Contents

Advertisement

3.3.2

Secretion canister

!
Important safety information on
the secretion canister system
1. Insert the charging plug of the charger into the
charging socket on the device.
2. Connect the battery recharging unit with the supplied
country specific power cable.
3. Plug the power plug into the power socket.
ª The ATMOS® C 051 Thorax now displays the icon
on the screen. The bar at the right flashes.As
long as the power plug is plugged in, the icon
is green.
When the battery is fully charged the icon
flashing.
4. Disconnect the power plug from the socket.
5. Remove the charging plug from the charging socket.
As soon as the battery charge level is less than 20%, the
drainage unit displays a warning window and triggers an
audible warning message (chapter „5.0 Warning messages"
on page 32) Charge the battery in order to continue the
therapy without interruption. If the battery is too low for
further operation the ATMOS® C 051 Thorax switches off
automatically.
The battery of the ATMOS® C 051 Thorax can also be
charged when the device is switched off. The state of
charge can be seen on the display.
• Always use the original ATMOS disposable secretion canister.
• Vacuum connection system: The vacuum connection between
device and secretion canister is set up immediately after
connection!
• For hygienic reasons we recommend an exchange of both
secretion canister and hose system at the same time.