Table of contents Introduction.......................4 Notes on operating instructions ................4 Explanation of pictures and symbols ..............6 Intended use .......................8 Function ........................9 Intended users ......................9 Scope of delivery ......................9 Transport and storage .................... 11 Notices for your safety ...................13 General safety instructions ..................13 Danger for users, patients, and third parties ............
Page 3
Consumables ....................35 Disposal ......................36 Technical data ....................37 Notes on EMC ....................39 Further information, accessories, consumables and spare parts are available from: ATMOS MedizinTechnik GmbH & Co. KG Ludwig-Kegel-Str. 16 79853 Lenzkirch Germany Phone +49 7653 689-0 +49 7653 689-190 (Switchboard) +49 7653 689-393 (Service) E-mail: info@atmosmed.de...
This document may only be reprint- ed, either in part or in whole, with written permission from ATMOS. These operating instructions must always be kept available near the product.
The Declarations of Conformity and our General Terms and Conditions can be viewed on our website at www.atmosmed.com. The quality management system at ATMOS has been certified accord- ing to the international standard EN ISO 13485. These operating instructions are valid for the following devices: • ATMOS C 451 340.0300.0 • ATMOS C 451 mobile with 2 x 1.5 l canisters – fully assembled 340.0376.0 • ATMOS C 451 mobile with 2 x 1.5 l canisters – for self-assembly 340.0335.0...
Page 6
DANGER Warning of a danger that will result in immediate fatal or serious injury. Observe the necessary measures. WARNING Warning of a danger that can cause fatal or serious injury. Observe the necessary measures. CAUTION Warning of a danger that can cause minor injury. Observe the necessary measures. ATTENTION Notice of a danger that can damage the product or other objects.
Page 7
European Article Number IPX1 Protection against the penetration of harmful moisture (dripping water) Application part type BF Professional disposal For single use only (symbol located on consumables) Not sterile STERILE Autoclavable AUTOCLAVE PATIENT Connection for suction hose / patient No natural rubber latex LATEX Potential equalisation Protection class II...
1.3 Intended use Name: ATMOS C 451 Main function: Suction of secretions, blood, serous fluids, rinsing fluids, and for the temporary collection of these fluids. Medical indications / For all applications where suction is needed, such as in gener- application: al surgical procedures (e.g., suction of wound cavities, ab- scesses), suction of the nasopharyngeal cavity, in endoscopy for suction of secretion or rinsing fluids, and in neurosurgery.
ATMOS C 451 Basic device Power cable 5 m 008.0818.0 Operating Instructions Name 340.0376.0/340.0335.0 ATMOS C 451 mobile with 2 x 1.5 l canisters 1x ATMOS C 451 340.0300.0 2x DDS secretion canister, plastic, 1.5 l, autoclavable 340.0050.0 2x DDS canister lid with sealings, autoclavable 340.0053.0 2x DDS canister handle, blue, autoclavable 340.0326.0...
Page 10
Name 340.0377.0/340.0334.0 ATMOS C 451 mobile with 2 x 3 l canisters 1x ATMOS C 451 340.0300.0 2x DDS secretion canister, plastic, 3 l, autoclavable 340.0051.0 2x DDS canister lid with sealings, autoclavable 340.0053.0 2x DDS canister handle, blue, autoclavable 340.0326.0 2x DDS splash protection, silicone, autoclavable 340.0056.0...
Only transport the product in a shipping carton that is padded and offers sufficient protection. If damage occurs during transport: 1. Document and report damages in transit. 2. Send the device to ATMOS; see chapter ‘7.3 Sending in the device’ on page 30. Ambient conditions for transport and storage: • Temperature: −30...+50 °C • Relative humidity: 5...90% without condensation...
Page 12
Storage: After transporting the device at temperatures below 0 °C, keep it at room tempera- ture for at least six hours before initial start-up. If the device has not been acclima- tised, it may not be operated as the electronic components may become damaged. 12 Introduction...
‘Intended use’ on page 8. • Never operate the product in potentially explosive areas or areas that are oxygenat- • Only use original accessories and original spare parts from ATMOS. This applies to the power cable in particular. WARNING Avoid misuse.
Page 14
• Always carry out a function check before using the device. • ATMOS recommends always having another suction device ready at hand. That way you can also perform suctioning if a device should fail. WARNING Risk of infection due to pathogens on the product! Deadly diseases may be transmitted.
Only use proper mains connections and extension cords. • Never touch the device's interfaces and the patient at the same time! • Only use original accessories and original spare parts from ATMOS. This applies to the power cable in particular. • Observe the specifications regarding periodic tests in chapter ‘7.1 Periodic tests’ on page 30.
3 Setting up and starting up Always place the device on a firm and level surface. 3.1 Device overview Front view DDS secretion canister lid DDS secretion canister Vacuum gauge On/off switch Vacuum regulator Vacuum connection: Direct Docking System ...
4 Operation 4.1 Initial start-up Observe the safety instructions prior to initial start-up! After transporting the device at low temperatures, it should be kept at room temper- ature for at least six hours before initial start-up; otherwise, the device may not be operated.
4.6 Inserting and removing the DDS bacterial and viral filter / oversuction stop The DDS bacterial and viral filter / oversuction stop is intended for single use only. Before each use, check that the DDS bacterial and viral filter / oversuction stop is dry and clean. Re- place the DDS bacterial and viral filter with a new DDS bacterial and viral filter if it is discoloured or contaminated or if oversuction has occurred. • Insert the DDS bacterial and viral filter onto the DDS secretion canister handle.
4.11 Connecting the suction hose Connect the suction hose to the already inserted hose adapter. 4.12 On/off switch • Press the “I” symbol to switch on the device. • Press the “O” symbol to switch off the device. • In position ‘O’ the device is ready for optional foot switch operation (see options).
3. Replace the DDS bacterial and viral filter with a new DDS bacterial and viral filter if it is discoloured or contaminated or if oversuction has occurred. 4. Switch on the device. 5. Close the suction hose and set the desired vacuum. 6. Connect the suction catheter, hose-rinsing aperture, or suction instruments. Observe the liquid level in the secretion canister while suctioning. The DDS bacterial and viral filter prevents liquid from being sucked into the pump. Nevertheless, the canister should be emptied or replaced when it is 2/3 full (including foam crown).
System trolley • For mobile use, a system trolley is available which can be used with two ATMOS secretion canister systems on the DDS changeover docking station or with disposable systems. • During operation always position the system trolley on a flat, stable surface.
5.1 Securing the device Safe operation of the mobile suction device is only guaranteed with the corresponding special system trolley or trolley with standard rail. • The suction device is placed on the system trolley in such a way that its feet engage in the holes of the trolley.
• The adapter with standard rail is required to use the bottom vacuum connection. This adapter seals the upper vacuum connection and allows you to hang accessories on the standard rail section now available. • Mount the adapter with standard rail to the mounting fixture of the suction device as de- scribed in the enclosed assembly instructions.
5.6 Use of disposable suction systems (Serres®, Medi- Vac®, Receptal®) Observe the operating instructions by the manufacturer of the corresponding canister system. Do not operate the device without a bacterial filter. Disposable systems for tabletop devices: Optionally, the suction device can also be used as a tabletop unit with disposable suction systems that can be attached to a standard rail.
6 Reprocessing 6.1 Safety instructions for reprocessing 6.1.1 General safety instructions We recommend that you always document all maintenance work and part replacements in writing. It is the responsibility of the user to ensure that the required results of cleaning and disinfection are achieved.
Unsuitable cleaning agents and disinfectants. The product may become damaged. • Do not use any process chemicals containing the following ingredients on plastic parts • Chloramides or phenol derivatives Do not use any process chemicals containing the following ingredientson stainless •...
Agent Active ingredients in 100 g Type Mikrobac® forte 19.9 g benzyl C12-18-alkyldimethylammonium chloride, 5 g N- (3-ami- Liquid (Bode Chemie) nopropyl) -N-dodecylpropane-1,3-diamine Concentrate Bacillol® 30 14 g ethanol, 10 g propan-2-ol, 6 g of propan-1-ol, 0.5 g N-alkyl-amino- Foam Foam (Bode propyl-glycine Ready-to-use...
Replace the DDS bacterial and viral filter at every cleaning or when disinfecting the DDS canister system. 6.3.2 Selecting process chemicals Always observe the manufacturer’s specifications for the process chemicals. Agent (manufacturer) Active ingredients in 100 g Type Cleaning agents – Mechanical reprocessing neodisher® <5% nonionic and anionic surfactants, Liquid MediClean forte enzymes concentrate (Dr. Weigert) neodisher® AN (Dr. <5% nonionic surfactants, >30% phos- Powder Weigert) phates, enzymes...
Page 29
Mechanical cleaning and Secure the accessories on a suitable load carrier. disinfection • Clean and disinfect using a suitable programme: Pre-wash: 1 min – Pre-rinse with cold water Clean: 5 min, 50 °C / 122 °F – Cleaning with a cleaning agent Neutralise: 2 min –...
7.1 Periodic tests • Perform a repeat test of the electrical safety according to IEC 62353 at least every 24 months. • In this context, ATMOS recommends arranging for an inspection in accordance with the manufacturer‘s specifications. 7.2 Function check • Prior to every use, perform a visual inspection of the device, hoses, secretion canis- ters, and connection cables.
7.5 Changing the fuse 1. Remove the power cable. 2. Press the two-sided spring clips of the fuse holder with a screwdriver and pull out the fuse holder. 3. Replace the fuse and insert the holder until both spring clips engage again. 4.
8 Troubleshooting The product has been subjected to a thorough quality control in the factory. If a fault should occur despite this care, you may be able to resolve it yourself. Error symptom Possible cause Remedy Device does not • Power plug fits poorly •...
9 Accessories Accessories Foot switch 443.0755.0 Foot controller (0...−91 kPa) 443.0310.1 Potential equalisation for the system trolley 340.0082.0 Potential equalisation cable 008.0596.0 System trolley ATMOS C 451 – for self-assembly 340.0070.0 System trolley ATMOS C 451 – fully assembled 340.0170.0 System trolley ATMOS C 451 with DDS changeover station – fully 340.0171.0 assembled System trolley ATMOS C 451 with standard rail – fully assembled 340.0172.0 Trolley with standard rail 320.0070.1 Standard rail set for system trolley 340.0081.0 Tray for system trolley 340.0084.0...
Page 34
Accessories Nipple set with overflow electrode 444.0012.0 Serres® outer canister, 1 l 312.0465.0 Serres® outer canister 2 l 310.0402.0 Serres® outer canister 3 l 310.0403.0 Standard rail holder Serres® complete 444.0484.0 Receptal® outer canister 1.5 l 310.0221.0 Receptal® outer canister 2 l 443.0256.0 Receptal® outer canister 3 l 444.0157.0 Receptal®...
The product does not contain any hazardous materials. 5. Clean and disinfect the device. 6. In Germany: Send the product back to ATMOS or your specialist dealer. They will dispose of the product properly. 7. In other countries: Dispose of the product properly and in accordance with coun- try-specific laws and regulations.
HF-shielded room of a magnetic resonance imaging system. The customer or user of the ATMOS C 451 must ensure that it is used in such a pre- scribed environment. Guidance and manufacturer’s declaration – key features ...