Reading Out The Therapy Data - Atmos C 051 Thorax Operating Instructions Manual

Table of Contents

Advertisement

If the therapy data should be transferred during a patients therapy, follow the steps below:
• Clamp the thoracic catheter
• Stop the current therapy
• Remove the secretion canister.
Perform the therapy data transfer as described.
• Connect the secretion canister.
• Continue the therapy
• Reopen the clamp at the thoracic catheter.
4.7.4

Reading out the therapy data

• Connect the USB flash drive to PC.
• Open the folder on the USB flash drive. This folder contains a PDF file and an excel file.
• Open the PDF file.
• Fill in the desired Information:
Patient data
1
Diagnosis
2
Description of the secretion
3
Following information can be seen in the report:
Beginning and end of recording, flow at beginning and end of recording
4
File name und device ID
5
Graphic diagram of the therapy data
6
Therapy Report
ATMOS® C 051 Thorax
Patient:
1
Diagnosis:
2
Secretion:
3
Start date:
12.02.2011 / 04:28
Air leak:
Flow @ Start 0.01 l/min
28
Complete data transfer
• As soon as the therapy data is transferred then the USB flash
drive may be removed. Now you return to the main screen.
End date:
Recording time:
12.02.2011 / 06:31
00:02:00:00
4
Flow @ End 0.01 l/min
File name:
ATMOS_C_051_THORAX_ID0C1C_20110212_063107.pdf
Device ID:
5
0C1C
6

Advertisement

Table of Contents
loading

Table of Contents