9.0 Maintenance and Service

9.1

Basic instructions

Maintenance, repairs and period tests may only be carried out by persons who have the
appropriate technical knowledge and are familiar with the product. To carry out these measures
the person must have the necessary test devices and original spare parts.
ATMOS recommends: Work should be carried out by an authorized ATMOS service partner. This
ensures that repairs and testing are carried out professionally, original spare parts are used and
warranty claims remain unaffected.
Carry out an inspection according to the manufacturer's specifications every 12 months. This
check should also include a check of the battery capacity as well as all parts for wear or damage.
The unit does not require any further maintenance.
A regular thorough cleaning and disinfection of the drainage unit and the application parts,
respectively the operation in line with the operating instructions are assumed.
Please observe any national and international regulations applicable for your institution.
9.2

Repairs

The following may require repairs from the manufacturer or an authorized service partner. Prior
to sending in the device, please contact your service partner by phone.
• Liquid has penetrated the drainage unit.
• Significant decrease of battery capacity.
• Sudden occurrence of abnormal displays on the screen.
• Sudden occurrence of unusual noises
• Operational and functional disorders which cannot be resolved by means of the hints
describes in the chapter „10.0 Troubleshooting" on page 48.
9.3

Sending in the device

If the drainage unit has to be sent in for repair after consultation with the manufacturer or an
authorized service partner, we ask you to observe the following:
1. Remove and properly dispose of consumables.
2. Clean and disinfect the product and accessories according to the operating instructions.
3. Place used accessories with the device.
4. Fill in the form QD 434 „Delivery complaint / return shipment" and the respective
decontamination certificate.
 This form is enclosed to each delivery and can be found at www.atmosmed.com.
5. The device must be well padded and packed in suitable packaging.
6. Place the form QD 434 „Delivery complaint / return shipment" and the respective
decontamination certificate in an envelope.
7. Affix the envelope to the outside of the package.
8. Send the product to ATMOS or to your dealer.