Press the <Right Arrow> and [Span Data] is highlighted.
Press the <Right Arrow> again and [Max LiquIdent] will appear. For example, to change
the [LiquIdent] data, press the <Right Arrow> and an equal sign appears. Using the Mag-
netic Wand or PC number keys, enter the new data (e.g., 1350). Press the <ENTER> key to
store the numeric data.
Controlotron
Channel
Set 100%
Max LiquIdent
Min Liquident
Max ROC
Min ROC
Max Vs m/s
Min Vs m/s
Max S.G.
Min S.G
Max API
Span Data
1.17

HOW TO SET UP A CHANNEL FOR TRANSIT-TIME OPERATION

This section shows how to set up a single-channel for a transit-time, clamp-on operation and
how to create a new site setup, enter pipe data and then select a liquid class (which tailors the
meter response to the liquid's density and viscosity). It explains how to select a clamp-on
transducer model, pick a mounting mode and, after physically mounting the transducers on
the pipe, invoke the [Install Transducer] procedure. This setup is the fastest way to get the
meter operational. It accepts most defaults and only provides a data output on the local LCD
display panel. The first step is to create a new site setup.
NOTE: Always begin a new installation by issuing the [Create/Name Site] command.
This action prepares the flow computer for site programming by loading sys-
tem defaults into the appropriate menu cells.

1.17.1 HOW TO ACCESS THE CHANNEL SETUP MENU AND CREATE A SITE

As shown is previously, when you access the Installation Menu, you will arrive at the top
menu location [Meter Type]. With single-channel meters, all you have to do is press the <Right
Arrow> twice to access the main section of the Clamp-On Menu (see next page).
Siemens Hardware
Controlotron
Channel
Span 0% and 100% Values for Analog Data
Span Data
Set Alarm Levels
Interface Alarms
Data Span/Set/Cal
[1]
SITE1
=0.00
0.00
0.00
0.00
0.00
0.00
0.00
0.00
0.00
1-31
[1]
SITE1
Controlotron
Channel
Set 100%
Max LiquIdent
Min Liquident
Max ROC
Min ROC
Max Vs m/s
Min Vs m/s
Max S.G.
Min S.G
Max API
Span Data
1010BXFM-3
[1]
SITE1
1350
0.00
0.00
0.00
0.00
0.00
0.00
0.00
0.00

Advertisement

Table of Contents
loading

Table of Contents