Section 2
2.9.3
PATH ENABLE (Not Applicable)
To select the Path Enable function:
To access [Path Enable] press
To select [Path Enable] press
To store the selection press
2.9.4

PATH DATA

The 1010 Interface Detector provides information and commands that can be used to optimize
performance, correct Site Setups, or other application problems. Although most commands
are self explanatory, the table below lists and explains the Path Data menu items.
Vs m/s
Signal (mV)
Valc %
Vaer %
Alarm Status
AnCal
Siemens Hardware
ENT

PATH DATA MENU ITEMS

Current liquid sonic velocity (Vs) in m/s. The Vs value depends on
the liquid and its temperature. This menu cell allows "fine-tuning"
when highlighted. Press the [+] key to increase, or the [-] key to
decrease the reading.
CAUTION: Incorrect usage will result in reduced accuracy.
Xdcr signal strength (in mV). This is the amplitude of the transit-
time receive signals. Improper coupling, attenuative liquid or pipe
could cause a low value.
Input Amplifier gain indication. Larger % indicates a stronger liquid
signal.
This shows the current percent of aeration detected by the meter.
This letter code field shows the status of the meter's built-in alarms.
A dash indicates an inactive alarm. Letter codes reflect the following
alarm conditions:
S = Transducer Spacing Warning
E = Empty Alarm
F = Fault Alarm
A = Aeration Alarm
M = Memory Activated (Fault Suppressed)
K = Makeup Flag (May be Latched)
I = Interface Alarm
Simulator for calibrating external devices, etc. Highlight this item,
then press the <Right Arrow> to enable numeric entry. Type desired
simulated API.
Note: AnCal remains active only while you remain in this menu
cell.