Datalogger Data Option List; Alarm Letter Codes And Descriptions - Siemens SITRANS F FUH1010 Series Operating Instructions Manual

Nema 7 compact interface detector
Table of Contents

Advertisement

Section 2
THE DATALOGGER SETUP MENU STRUCTURE (continued)
Datalogger Setup
None
Default – Datalogging is disabled.
Site ID
You entered this site name when you created the site.
Date
The report date according to the meter's real time clock. (Format: MM.DD.YYYY)
Time
The report time according to the meter's real time clock. (Format: HH.MM)*
Path Vs
The sonic velocity of the liquid in either Path 1 or Path 2.
V s
The average liquid sonic velocity since the last report.
Path Valc
The signal amplitude of the liquid in either Path 1 or Path 2.
Valc
The signal strength measured at the time of the report.
Aeration
The aeration percentage recorded at the time of the report.
Path Alarms
The letter codes of any Path related alarms at the time of the report.
Alarms
The letter codes of any alarms active at the time of the report (see below).
S.G
Specific Gravity at current operating temperature.
Base S.G.
Specific Gravity referenced to user selected temperature.
API#
API number at current operating temperature.
Base API#
API number referenced to user selectable temperature.
Kg/m3
Kilograms per cubic meter at current operating temperature.
Base Kg/m3
Kilograms per cubic meter referenced to user selectable temperature.
Temperature
The current instantaneous liquid temperature.
Delta T (uSecs)
The Up to Down transit-time difference measured at the time of the report.
Analog Inputs
Allows the device to accept analog data from external devices (i.e., pressure, temperature)
Sonilocate
Detects a sudden drop in pipe pressure. The Vs decrease is detected and time-stamped.
*The Time field expands to HH.MM.SS when you select a logging interval of more than once per minute.
If you select Alarms, the Datalogger logs the state of each alarm function upon generating
a report. A dash [-] represents an inactive alarm. A letter code represents an active alarm:

Alarm Letter Codes and Descriptions

S ....... Spacing ............... Transducer spacing may need re-adjustment.
E ...... Empty ................ Pipe is empty.
F ...... Fault .................. Three continuous seconds without new data update.
A ...... Aeration ............ Current aeration percentage exceeds the alarm setpoint.
M ..... Memory ............. Last valid reading for a selected interval during Fault condition.
K ...... Makeup ............. In-Process Makeup occurred.
I ....... Interface ............ Liquid Vs exceeds interface alarm setpoint.
LogTime Interval
Datalog Events

DATALOGGER DATA OPTION LIST

2-44
1010BXFM-3
4 Hr.
6 Hr.
12 Hr.
24 Hr.
Demand
None
Relay 2
Relay 1
Spacing
Empty
Aeration
ROC
Fault Alarm
Memory
Makeup