FUH1010 Interface Detector Compact NEMA 7 Manual Updates
1. FUH1010 Interface Detector NEMA 7 Compact Flow Meter Manual
Troubleshooting Table Update
In the FUH1010 Interface Detector NEMA 7 Compact flow meter manual P/N CQO:1010BXFM-3 Revison
B replace the troubleshooting table in Section 2 - Installation Menu on pages 2-58, 2-59 and 2-60 with
the following:
Error or Message
Memory Full!
Memory Corrupted!
Chan Not Setup
Clr Active Memory?
Clr Save Data?
<EOT>
No Sites - Press <ENTER>
Security
RTC Error
Siemens Hardware
ERRATA
Probable Cause
Response to an attempt to save site
data when data memory is full.
Memory read error occurred while
accessing the active site data.
Response to an attempt to invoke
an operation that requires a chan-
nel to be enabled.
Use this function to restore opera-
tion if a severe event (e.g. a violent
power surge) disrupts system
operation.
[Clr Saved Data?] only appears
after pressing <Down Arrow> in
response to [Clr Active Memory?].
Response to a request to output
Datalogger data to the printer or
the Graphics screen when no
datalogger data exists or at the end
of a transmitted file.
Response while trying to recall/
delete a site setup when no sites
are stored.
Response upon charging previously
entered data when security switch
is in [Disable] position or security
code has been entered.
Component level problem.
E-1
1010FMA-61
Solution
Delete an obsolete site or clear
Datalogger memory to make
room for the new data.
Refer to F4 reset procedure in
the Operation Instructions
manual.
Enable the channel [Channel
Setup Channel - Enable - Yes].
Note that a channel cannot be
enabled until an "Install" opera-
tion is completed.
Refer to F4 reset procedure in
the Operation Instructions
manual.
Answering [Yes] to [Clr Saved
Data?] will erase ALL saved
data. To invoke in RS-232 serial
mode, type @@@ and then
press <ENTER> key.
Set up the Datalogger.
Create a site.
Change switch position to
[Enable].
Enter previously set security
code.
Meter requires service. Request
RMA.
(continued)