POWER CONNECTIONS

1.7.1 INPUT POWER CONNECTOR (P8) AND FUSE (F1) LOCATIONS

The 1010BX Interface Detector requires 100/120/200 VAC or 9-36 VDC for its power input.
The power input is protected by a GMA 2A fuse (F1). Please replace this fuse with the exact
type if it becomes necessary to do so.
This must be
connected to Earth
Ground.

1.7.2 POWER SUPPLY WIRING

NOTE: Refer to installation drawings 1010BX-7 (1 of 3) and 1010BDX-7 (1 of 3).
WARNING: Turn off main power before installing AC connections to the 1010 NEMA
flowmeter. Contact with exposed wiring may lead to fire, electric shock,
or serious personal injury.
Using a 1/16" Hex key, loosen the 1010X Rear Housing Cover locking setscrew.
Unscrew the Rear Housing Cover and remove.
Locate power supply connector J8.
Pull wires through flowmeter case conduit cable hole.
Wire input power connector P8 for AC or DC power as shown below. Insert wires into
wire entry holes and secure by tightening wire clamp screws.
NOTE: Power Supply connector wires should be stripped stranded or solid conductors
Siemens Hardware
AWG 12 -18.
Fuse (F1) GMA 2A
Cable Clamp
1-5
1010BXFM-3
Loosen Wire Clamp
Screw, insert stripped
end of wire and then
re-tighten.
Strip wire insulation
back by 0.031" (8mm).