However, assume that the liquid's sonic velocity is very well known and is actually 1377
m/s. Press the <Right Arrow>. Note that an "equal sign" appears before the displayed
sonic velocity. Type [1377] and then press the [ENTER] key.
NOTE: As your 1010BX metering experience with various liquids evolves, your knowl-
edge and confidence regarding their sonic velocities will grow. If the Estimated
Vs and Measured Vs at site are different, as in the example on the previous
page, consider it a warning that something about the installation may be incor-
rect. Recheck transducer mounting geometry (e.g., Are they axial, diametrical
and spaced correctly?) and your pipe data and re-install. On the other hand,
when the Measured Vs conforms to the Actual Vs, your degree of confidence in
the correctness of the installation and the meter's intrinsic calibration perfor-
mance should be high.
This commands the meter to re-calculate its timing parameters to accommodate the new
sonic velocity setting.
Siemens Hardware
Controlotron
2 Channel
Drive 53
0
36
Transducer Model
Transducer Size
Xdcr Mount Mode
Measured Vs m/s
Spacing Offset
1395
Number Index
Spacing Method
Ltn Value
Install Complete?
Empty Pipe Set
Pick/Install Xdcr
Controlotron
2 Channel
Drive 53
0
36
Transducer Model
Transducer Size
Xdcr Mount Mode
Measured Vs m/s
Spacing Offset
1377
Number Index
Spacing Method
Ltn Value
Install Complete?
Empty Pipe Set
Pick/Install Xdcr
1-37
[1]
SITE1
[6:
:0]
1011H Hi Prec
C1H
Reflect
Spacer Bar 1012BP
2.330
Install
Chan Not Setup
[1]
SITE1
[6:
:0]
1011H Hi Prec
C1H
Reflect
Spacer Bar 1012BP
2.330
Install
Chan Not Setup