System Reset Using The Magnetic Wand - Siemens SITRANS F FUH1010 Series Operating Instructions Manual

Nema 7 compact interface detector
Table of Contents

Advertisement

Section 1

1.14.1 SYSTEM RESET USING THE MAGNETIC WAND

As mentioned previously, there is a power-on delay of approximately 30 seconds from the
time that you apply power to the 1010BX Interface Detector to when the 16 x 2 Character LCD
becomes active. The reason for the delay is to give you a chance to perform the following
system reset procedure.
To clear Active Memory using the Magnetic Wand:
Turn off power (if it is currently on) and the reapply power to the system.
Trigger
during the power-on delay. Examine the left-hand figure below and note the
message on the LCD display:
Clr Active Memory?
>No
Trigger
. Note that the greater than (>) symbol changes to a (:) colon. Trigger
change the second line to [Yes] (as shown in the right-hand figure above).
Trigger
to clear the contents of Active Memory (but preserve any stored Site Set-
ups).
To restore operation, create a new site setup or recall a stored site setup and then re-
select any Meter Facilities items (e.g., RS-232 setup parameters) that you may have
edited originally.
To clear ALL Dynamic Memory using the Magnetic Wand:
CAUTION: Before proceeding further it is essential to understand that this function
eliminates ALL data stored in RAM. This means that all saved Site Setups
including the site data of a flow-calibrated site will be erased! In addition,
the entire Datalogger file plus any custom factory or user-created pipe or
transducer tables will be eliminated. The impact of this is such that we
strongly recommend that you consult the Technical Service Department
before continuing with this procedure. If you choose to continue, be aware
that you will have to create a new Site Setup, re-enter all site specific pa-
rameters including pipe or transducer tables, plus all desired Meter Fa-
cilities entries.
Turn off power (if it is currently on) and then reapply power to the system.
Trigger
during the power-on delay. Trigger
Clr Active Memory?
:Yes
or
. Clr Saved Data? will appear.
1-24
1010BXFM-3
to

Advertisement

Table of Contents
loading

Table of Contents