4.

THE METER FACILITIES MENU

Meter Facilities functions are available immediately after pressing
menu provides global control that enables you to identify and activate the following functions
and features of the hardware supported by this model.
Select data entry and meter output in either English (default) or metric units.
Customize the default pipe and transducer tables to suit your requirements.
Output, display, erase and select a memory management method for the Datalogger.
Determine the amount of memory available.
Verify/Adjust the analog output voltage and current using a multimeter.
Verify/Adjust the analog pulse output using a frequency counter.
Calibrate the RTD temperature sensors.
Set the system clock/calendar.
Obtain detailed software/hardware identification.
This menu allows you
to specify default
measurement units.
This menu allows you
to create or edit pipe
and transducer tables.
This menu allows you
to work with and
select a Datalogger
storage method.
This menu allows you to
calibrate the RTD
Temperature Sensors.
4.1

PREFERRED UNITS

The selection you make in this menu cell becomes the default units for all menu items and
data outputs. You can choose either English or metric units. The factory setting is English.
To specify metric default units for all meter functions:
To access the [Preferred Units] option list press
Press
To register your selection press
Siemens Hardware
Controlotron
Select Default Units for all Menus
Preferred Units
Table Setups
Datalogger Cntrl
Memory Control
Analog Output Trim
RTD Calibrate
Clock Set
RS-232 Setup
System Info
Meter Facilities
This menu shows detailed information
about this model's hardware and software.
to move the cursor to [Metric] menu cell.
ENT
Dual Path
SITE 1
English
09.02.99 09.30.06
9600 Odd 7
4-1
1010BXFM-3
. The Meter Facilities
MENU
This menu allows