Specification
Display Chart Speeds
Messages Screen
CE Conformity (CE Mark)
Immunity Product Classification
Enclosure Rating
Installation Requirements
EMC Standards
Safety
Specification
Number of Inputs
Input Types
Minimum Input Span
Burnout (T/C)
Cold Junction Compensation
Input Resolution
Input Impedance
Source Impedance
Square Root Extraction
Sensor Compensation
Input Sampling Rate
43-TV-25-35 GLO Iss.4 Dec 06 UK
A5E01001767-04
Specification Tables
Chart rates: 1 mm/hour, 5 mm/hour, 10 mm/hour, 20 mm/hour, 30 mm/hour, 60 mm/hour, 120 mm/
hour, 600 mm/hour, 1200 mm/hour, 6000 mm/hour. Combinations of rates can be mixed and chart
speeds can be set independently for each chart. Display speeds are independent of logging rate
The message screen displays system information and records any setup activity that has been
changed. It also provides warning and error message updates, lists alarm activity and will display
user defined marks on a chart.
This product conforms with the protection requirements of the following European Council Direc-
tives: 73/23/EEC, the Low Voltage Directive, and 89/336/EEC, the EMC Directive. Conformity of
this product with any other "CE Mark" Directive(s) shall not be assumed.
Complies with EN61326 Class I: Cord Connected, Panel Mounted Industrial Control Equipment
with protective earthing (grounding). (EN 61010-1)
Front panel designed to NEMA3 / IP54 (Optional NEMA 4X / IP66)
Category II: Overvoltage (EN 61010-1)
Pollution Degree 2
Emissions - EN61326 Class B (D200: Class A radiated)
Immunity - EN61326 Industrial Levels (D200: Light Industrial)
Complies with EN61010-1: 2001. Panel Mounted Equipment, Terminals must be enclosed within
the panel.
Analogue Inputs
D300 recorder can have 4, 6, 8, 12 or 16 input channels. D400 recorder can have 4, 6, 8, 12, 16,
24, 32, 40 or 48 input channels. D200 recorder can have 3, 6, 9 or 12 input channels.
mV, V, mA with external shunt (provided as standard), Thermocouple, RTD and ohms
Range is fully configurable with span limitation of the operating range selected with 4% under
range to 4% over-range capability (50V Range 2%)
Active (High or Low), Passive, and Health watch/Maintenance (option).
Internal compensation with the ability to manually adjust values, External Input for compensation,
External CJC value specified
0.0015% (16 Bit ADC)
Current loop resistance: 10 ohms, use ±0.1% external resistor. Volts >1M
T/C and RTD: 100 ohms per lead maximum (CU10 = 15 ohms, not available for the D200) A
single point cal on Slot A of the D200 recorder will improve accuracy for lead resistances above
10 ohms
Available as standard on every input type
Single point and Dual point
D300/D200 Recorder has 2 available slots and the D400 Recorder has 6 available slots with up
to 8 analog inputs each; the input sampling rate is dependent on actuation type.
D300/D400: All Inputs: 100mS (10Hz), 200mS (5Hz), 500mS (2Hz)
D300/D400: Fast Sampling: 20mS (50Hz) - mA, mV, Ohms and Volts only
D200: 200mS (5Hz), 500mS (2Hz) and 100mS (10Hz) (Analogue Input expansion card option)
Design Attributes
Ω
Ω
, all other >10M
225