Specification
Data Storage
Power Requirements
Common Relay Output (SPNC)
Battery
Password Protection
Languages
Temperature Units
Recorder Identification
Clock
Alarm Set Points
Data Replay Mode
224
Specification Tables
Removable Media: Compact Flash card supports up to 4.0 Giga bytes. (Compact Flash is not avail-
able as an option for the D200)
Local Mass Storage Options: USB memory key - up to 2Gb, USB hard drive - up to 120Gb,
Internal Data Buffer D300 and D400 Recorder: Non-volatile. 70MB (16 million acquisition values)
upwards to 1850MB (up to 400 million points)
Internal Data Buffer D200 Recorder: Non-volatile. 70MB (16 million acquisition values) upwards to
400MB (up to 90 million points)
Setup and screens: Stored internally on non-volatile memory
Manual Saving: Data saving by inserting Compact Flash card or USB memory stick
Data Saving Period: Related to log rate, number of pens, totals and alarms. Each pen is capable of
its own independent storage rate.
D300 and D400 Recorder: 20ms to 60h,
D200 Recorder: 100mS to 60h.
Data Format: binary encoded format
Recycling Mode: Internal memory has FIFO (First In First Out) capability where the newest data
over-writes the oldest data.
D300 Recorder
Voltage (VRMS): 100VAC to 250VAC (auto select). Frequency: 50/60Hz
Power Consumption: <40W.
24V Optional instrument power: 20 to 55VDC/20 to 30VAC. Power Consumption: < 40 watts
D400 Recorder
Voltage (VRMS): 100VAC to 250VAC (auto select). Frequency: 50/60Hz
Power Consumption: <60W.
24V Optional instrument power: 20 to 55VDC/20 to 30VAC. Power Consumption: < 60 watts
D200 Recorder
Voltage (VRMS): 100VAC to 250VAC (auto select). Frequency: 50/60Hz
Power Consumption: <40W.
24V Optional instrument power 20 to 30VDC/20 to 25VAC. Power Consumption: < 40 watts
NC common alarm relay: Two contacts, normally open when the recorder is powered (no active
alarms). Rating 24V, 1 Amp. (Not available for the D200)
Battery backed up for clock, replaceable Lithium battery Type 6032, 3.0V – 10 years life (Recorder
powered),
4 years life, typical (Recorder unpowered).
Multiple Administrator control of password setup and management with four levels of password
protection for – Engineer, Supervisor, Technician, and Operator. Up to 50 different users are avail-
able. Password protection restricts user entry to the recorder set up and specific screens.
Engineer – Highest access to all levels, Supervisor, Technician and Operator.
Supervisor – 2nd highest level including Technician and Operator access
Technician – 3rd level including Operator access
Operator – 4th and lowest level of access.
English UK & US, French, German, Italian, Spanish, Brazilian, Polish, Hungarian, Slovakian,
Czech, Turkish, Romanian, Russian, Portuguese and Greek.
°C, °F, °K
Status bar: Alternately displays Recorder ID and Recorder Screen Name. Displays Time and Date.
Accuracy: ±29ppm (±1 minute/month) @ 25°C.
Summer/Winter manual or automatic time adjustment or via communications. SNTP Client and/or
Server included for synchronising over Ethernet.
6 per pen integral "soft" alarm set points easily set by user to announce selected out of limit condi-
tions; user can select if an alarm triggers a change in the screen background colour. Alarm triggers
can be set for Hi, Lo, Deviation (latched or unlatched) for alarm acknowledgement.
Alarm Damping – 1 sec to 24 Hours; Hysteresis - +/- 100% of pen scale
Common relay output: 1A 24V, can be activated on any alarm. (Not available for the D200)
Data replay facility on chart displays at normal, fast or slow speeds with zoom and cursor.
Design Attributes
43-TV-25-35 GLO Iss.4 Dec 06 UK
A5E01001767-04