Measured variable
Measuring range
8
Заказывайте на сайте: www.tinnova.ru/endress_hauser/ || Эл. почта: info@tinnova.ru
Input
Direct measured variables
• Mass flow
• Density
• Temperature
Calculated measured variables
• Volume flow
• Corrected volume flow
• Reference density
Measuring ranges for liquids
DN
[mm]
8
15
25
40
50
80
Measuring ranges for gases
The full scale values depend on the density of the gas and can be calculated with the formula below:
=
· ρ
: x
max(G)
max(F)
G
Maximum full scale value for gas [kg/h]
max(G)
Maximum full scale value for liquid [kg/h]
max(F)
<
max(G)
max(F)
ρ
Gas density in [kg/m³] at operating conditions
G
x
Constant dependent on nominal diameter
DN
[mm]
8
15
25
40
50
80
To calculate the measuring range, use the Applicator sizing tool → 88
Calculation example for gas
• Sensor: Promass E, DN 50
• Gas: Air with a density of 60.3 kg/m³ (at 20 °C and 50 bar)
• Measuring range (liquid): 70 000 kg/h
• x = 125 kg/m³ (for Promass E, DN 50)
Measuring range full scale values
[in]
³⁄₈
0 to 2 000
½
0 to 6 500
1
0 to 18 000
1½
0 to 45 000
2
0 to 70 000
3
0 to 180 000
can never be greater than
max(G)
[in]
³⁄₈
½
1
1½
2
3
Proline Promass E 100
min(F)
[kg/h]
[lb/min]
0 to 73.50
0 to 238.9
0 to 661.5
0 to 1 654
0 to 2 573
0 to 6 615
max(F)
x
[kg/m
3
]
85
110
125
125
125
155
Endress+Hauser
to
max(F)