Mounting Safety Barrier
Promass 100
Ambient temperature range
Storage temperature
Climate class
Degree of protection
Vibration resistance
Shock resistance
Impact resistance
48
Заказывайте на сайте: www.tinnova.ru/endress_hauser/ || Эл. почта: info@tinnova.ru
23
Minimum distance between additional Safety Barrier Promass 100 or other modules.Engineering unit mm
(in)
Environment
Measuring device
Safety Barrier Promass 100
‣
If operating outdoors:
Avoid direct sunlight, particularly in warm climatic regions.
–40 to +80 °C (–40 to +176 °F), preferably at +20 °C (+68 °F)
DIN EN 60068-2-38 (test Z/AD)
Transmitter and sensor
• As standard: IP66/67, type 4X enclosure
• With the order code for "Sensor options", option CM: IP69 can also be ordered
• When housing is open: IP20, type 1 enclosure
• Display module: IP20, type 1 enclosure
Safety Barrier Promass 100
IP20
• Vibration, sinusoidal according to IEC 60068-2-6
– 2 to 8.4 Hz, 3.5 mm peak
– 8.4 to 2 000 Hz, 1 g peak
• Vibration broad-band random, according to IEC 60068-2-64
– 10 to 200 Hz, 0.003 g
2
/Hz
– 200 to 2 000 Hz, 0.001 g
– Total: 1.54 g rms
Shock, half-sine according to IEC 60068-2-27
6 ms 30 g
Rough handling shocks according to IEC 60068-2-31
> 22.5
> 22.5
(> 0.89)
(> 0.89)
• –40 to +60 °C (–40 to +140 °F)
• Order code for "Test, certificate", option JM:
–50 to +60 °C (–58 to +140 °F)
–40 to +60 °C (–40 to +140 °F)
2
/Hz
Proline Promass E 100
A0016894
Endress+Hauser