Pin Assignment, Device Plug - Endress+Hauser Proline Promass E 100 Manual

Coriolis flowmeter
Hide thumbs Also See for Proline Promass E 100:
Table of Contents

Advertisement

Pin assignment, device plug

30
Заказывайте на сайте: www.tinnova.ru/endress_hauser/ || Эл. почта: info@tinnova.ru
Safety Barrier Promass 100
1
A
2
 8
Safety Barrier Promass 100 with terminals
1
Non-hazardous area, Zone 2, Class I Division 2
2
Intrinsically safe area
Order codes for the M12x1 connectors, see the "Order code for electrical connection" column:
• 4-20 mA HART, pulse/frequency/switch output →  23
• PROFIBUS DP→  25
• Modbus RS485 →  26
• EtherNet/IP →  28
• PROFINET→  29
Supply voltage
For all connection versions except MODBUS RS485 intrinsically safe (device side)
Device plug MODBUS RS485 intrinsically safe with supply voltage →  31
Pin
2
1
2
1
3
3
5
4
4
5
A0016809
Coding
The following is recommended as a socket:
• Binder, series 763, part no. 79 3440 35 05
• Alternatively: Phoenix part no. 1669767 SAC-5P-M12MS
– With the order code for "Output", option B: 4-20 mA HART, pulse/frequency/switch output
– With the order code for "Output", option N: EtherNet/IP
• When using the device in a hazardous location: Use a suitably certified socket.
L+
L-
A
Proline Promass E 100
2
1
26
27
L
L+
B
A
Power
Modbus
supply
RS485
24V
DC
Safe area
Power
Lift panel for
bus termination
Communication
A
Safety Barrier
Promass 100
Hazardous area
Power
Modbus
supply
RS485
L
L+
B
A
20
10
62
72
Assignment
DC 24 V
Not assigned
Not assigned
DC 24 V
Grounding/shielding
Plug/socket
Plug
Endress+Hauser
A0030220

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents