Proline Promass E 100
Endress+Hauser
Заказывайте на сайте: www.tinnova.ru/endress_hauser/ || Эл. почта: info@tinnova.ru
PROFIBUS DP connection version
For use in the non-hazardous area and Zone 2/Div. 2
Order code for "Output", option L
Depending on the housing version, the transmitters can be ordered with terminals or device plugs.
Connection methods available
Order code
"Housing"
Output
Options
Terminals
A, B
Options
Device plugs
A, B
→  31
Options
Device plugs
A, B, C
→  31
Order code for "Housing":
• Option A: compact, coated aluminum
• Option B: compact, hygienic, stainless
• Option C ultra-compact, hygienic, stainless
 3
PROFIBUS DP terminal assignment
1
Power supply: DC 24 V
2
PROFIBUS DP
Order code
"Output"
Option L
Order code for "Output":
Option L: PROFIBUS DP, for use in non-hazardous areas and Zone 2/Div. 2
Possible options for order code
Power
"Electrical connection"
supply
Terminals
• Option A: coupling M20x1
• Option B: thread M20x1
• Option C: thread G ½"
• Option D: thread NPT ½"
Terminals
• Option L: plug M12x1 + thread NPT ½"
• Option N: plug M12x1 + coupling M20
• Option P: plug M12x1 + thread G ½"
• Option U: plug M12x1 + thread M20
Device plugs
Option Q: 2 x plug M12x1
→  31
26
B
27
A
1
L
+
_
2
L
Terminal number
Power supply
2 (L-)
1 (L+)
26 (RxD/TxD-P)
DC 24 V
2
1
A0022716
Output
27 (RxD/TxD-
N)
B
A
25