Depending on the housing version, the transmitters can be ordered with terminals or device plugs.
Connection methods available
Order code
"Housing"
Outputs
Options
Terminals
A, B
Options
Device plugs
A, B
→  31
Options
Device plugs
A, B, C
→  31
Order code for "Housing":
• Option A: compact, coated aluminum
• Option B: compact, hygienic, stainless
• Option C ultra-compact, hygienic, stainless
 2
Terminal assignment 4-20 mA HART with pulse/frequency/switch output
1
Power supply: DC 24 V
2
Output 1: 4-20 mA HART (active)
3
Output 2: pulse/frequency/switch output (passive)
Order code
"Output"
2 (L-)
Option B
Order code for "Output":
Option B: 4-20 mA HART with pulse/frequency/switch output
Power
supply
Terminals
• Option A: coupling M20x1
• Option B: thread M20x1
• Option C: thread G ½"
• Option D: thread NPT ½"
Terminals
• Option L: plug M12x1 + thread NPT ½"
• Option N: plug M12x1 + coupling M20
• Option P: plug M12x1 + thread G ½"
• Option U: plug M12x1 + thread M20
Device plugs
Option Q: 2 x plug M12x1
→  31
+
24
_
25
+
26
_
27
1
+
L
_
2
L
Terminal number
Power supply
Output 1
1 (L+)
27 (–)
DC 24 V
4-20 mA HART (active)
Proline Promass E 100
Possible options for order code
"Electrical connection"
3
2
1
Output 2
26 (+)
25 (–)
24 (+)
Pulse/frequency/switch
output (passive)
Endress+Hauser
A0016888