Interior Cleaning; Electromagnetic Compatibility (Emc); Process; Medium Temperature Range - Endress+Hauser Proline Promass E 100 Manual

Coriolis flowmeter
Hide thumbs Also See for Proline Promass E 100:
Table of Contents

Advertisement

Proline Promass E 100

Interior cleaning

Electromagnetic
compatibility (EMC)

Medium temperature range

Density

Endress+Hauser
Заказывайте на сайте: www.tinnova.ru/endress_hauser/ || Эл. почта: info@tinnova.ru
• Cleaning in place (CIP)
• Sterilization in place (SIP)
Options
Oil- and grease-free version for wetted parts, without declaration
Order code for "Service", option HA
• Depends on the communication protocol:
– HART, PROFIBUS DP, EtherNet/IP:
As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
– Modbus RS485:
As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
– PROFINET: as per IEC/EN 61326
• Complies with emission limits for industry as per EN 55011 (Class A)
• Device version with PROFIBUS DP: Complies with emission limits for industry as per EN 50170
Volume 2, IEC 61784
The following applies for PROFIBUS DP: If baud rates > 1.5 MBaud, an EMC cable entry must be
used and the cable shield must continue as far as the terminal wherever possible.
Details are provided in the Declaration of Conformity.

Process

–40 to +150 °C (–40 to +302 °F)
Dependency of ambient temperature on medium temperature
T
a
 24
Exemplary representation, values in the table below.
T
Ambient temperature
a
T
Medium temperature
m
A
Maximum permitted medium temperature T
require a reduced ambient temperature T
B
Maximum permitted ambient temperature T
sensor
Values for devices used in the hazardous area:
Separate Ex documentation (XA) for the device .
0 to 5 000 kg/m
3
(0 to 312 lb/cf)
A
B
at T
= 60 °C (140 °F); higher medium temperatures T
m
a max
a
for the maximum specified medium temperature T
a
T
m
A0031121
m
of the
m
49