Electrical connection

34
Заказывайте на сайте: www.tinnova.ru/endress_hauser/ || Эл. почта: info@tinnova.ru
Connecting the transmitter
A
1
A
Housing version: compact, aluminum coated
B
Housing version: compact hygienic, stainless
1
Cable entry or device plug for signal transmission
2
Cable entry or device plug for supply voltage
C
Housing version: ultra-compact, hygienic, stainless, M12 device plug
3
Device plug for signal transmission
4
Device plug for supply voltage
• Terminal assignment→  23
• Pin assignment, device plug→  30
In the case of device versions with a connector, the transmitter housing does not need to be
opened to connect the signal cable or power supply cable.
Connection examples
Current output 4 to 20 mA HART
1
2
 9
Connection example for 4 to 20 mA HART current output (active)
1
Automation system with current input (e.g. PLC)
2
Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
specifications →  39
3
Connection for HART operating devices →  79
4
Resistor for HART communication (≥ 250 Ω): observe maximum load
5
Analog display unit: observe maximum load
6
Transmitter
B
2
1
2
3
4...20 mA
4
5
Proline Promass E 100
C
3
4
6
Endress+Hauser
A0016924
A0029055