Low Flow Cut Off; Protocol-Specific Data - Endress+Hauser Proline Promass E 100 Manual

Coriolis flowmeter
Hide thumbs Also See for Proline Promass E 100:
Table of Contents

Advertisement

Proline Promass E 100

Low flow cut off

Protocol-specific data

Endress+Hauser
Заказывайте на сайте: www.tinnova.ru/endress_hauser/ || Эл. почта: info@tinnova.ru
Intrinsically safe values
Supply voltage
20 (L-)
With IIC
With IIB: L
For an overview and for information on the interdependencies between the gas group - sensor - nominal
diameter, see the "Safety Instructions" (XA) document for the measuring device
1)
The gas group depends on the sensor and nominal diameter ff.
Transmitter
Intrinsically safe values
Order code for
"Approval"
• Option BM: ATEX II2G + IECEx Z1 Ex ia, II2D Ex tb
• Option BO: ATEX II1/2G + IECEx Z0/Z1 Ex ia, II2D
• Option BQ: ATEX II1/2G + IECEx Z0/Z1 Ex ia
• Option BU: ATEX II2G + IECEx Z1 Ex ia
• Option C2: CSA C/US IS Cl. I, II, III Div. 1
• Option 85: ATEX II2G + IECEx Z1 Ex ia + CSA C/US
IS Cl. I, II, III Div. 1
For an overview and for information on the interdependencies between the gas group - sensor - nominal
diameter, see the "Safety Instructions" (XA) document for the measuring device
The switch points for low flow cut off are user-selectable.
HART
Manufacturer ID
Device type ID
HART protocol revision
Device description files
(DTM, DD)
HART load
Terminal numbers
10 (L+)
U
= 16.24 V
o
I
= 623 mA
o
P
= 2.45 W
o
1)
: L
= 92.8 µH, C
= 0.433 μF, L
o
o
= 372 µH, C
= 2.57 μF, L
o
o
o
Supply voltage
20 (L-)
0x11
0x4A
7
Information and files under:
www.endress.com
Min. 250 Ω
Signal transmission
62 (A)
72 (B)
/R
= 14.6 μH/Ω
o
o
/R
= 58.3 μH/Ω
o
Terminal numbers
Signal transmission
10 (L+)
62 (A)
U
= 16.24 V
i
I
= 623 mA
i
P
= 2.45 W
i
L
= 0 µH
i
C
= 6 nF
i
72 (B)
13