Performance Characteristics; Reference Operating Conditions; Maximum Measured Error - Endress+Hauser Proline Promass E 100 Manual

Coriolis flowmeter
Hide thumbs Also See for Proline Promass E 100:
Table of Contents

Advertisement

Proline Promass E 100
Reference operating
conditions

Maximum measured error

Endress+Hauser
Заказывайте на сайте: www.tinnova.ru/endress_hauser/ || Эл. почта: info@tinnova.ru
The maximum cable length for individual wire cross-sections is specified in the table below. Observe
the maximum capacitance and inductance per unit length of the cable and connection values for
hazardous areas .
Wire cross-section
[mm
2
]
0.5
0.75
1.0
1.5
2.5

Performance characteristics

• Error limits based on ISO 11631
• Water with +15 to +45 °C (+59 to +113 °F) at2 to 6 bar (29 to 87 psi)
• Specifications as per calibration protocol
• Accuracy based on accredited calibration rigs that are traced to ISO 17025.
To obtain measured errors, use the Applicator sizing tool →  88
o.r. = of reading; 1 g/cm³ = 1 kg/l; T = medium temperature
Base accuracy
Design fundamentals →  44
Mass flow and volume flow (liquids)
• ±0.15 % o.r.
±0.10 % o.r. (order code for "Calibration flow", option A, B, C, for mass flow)
• ±0.25 % o.r.
Mass flow (gases)
±0.35 % o.r.
Density (liquids)
Under

reference operating conditions

[g/cm³]
±0.0005
1)
Valid over the entire temperature and density range
Temperature
±0.5 °C ± 0.005 · T °C (±0.9 °F ± 0.003 · (T – 32) °F)
[AWG]
[m]
20
70
18
100
17
100
16
200
14
300
Maximum cable length
[ft]
230
328
328
656
984
Standard density
1)
calibration
[g/cm³]
±0.02
41