Endress+Hauser Proline Promass E 100 Manual page 42

Coriolis flowmeter
Hide thumbs Also See for Proline Promass E 100:
Table of Contents

Advertisement

42
Заказывайте на сайте: www.tinnova.ru/endress_hauser/ || Эл. почта: info@tinnova.ru
Zero point stability
DN
[mm]
8
15
25
40
50
80
Flow values
Flow values as turndown parameter depending on nominal diameter.
SI units
DN
1:1
[mm]
[kg/h]
8
2 000
15
6 500
25
18 000
40
45 000
50
70 000
80
180 000
US units
DN
1:1
[inch]
[lb/min]
³⁄₈
73.50
½
238.9
1
661.5
1 654
2
2 573
3
6 615
Accuracy of outputs
The output accuracy must be factored into the measured error if analog outputs are used, but
can be ignored for fieldbus outputs (e.g. Modbus RS485, EtherNet/IP).
The outputs have the following base accuracy specifications.
Current output
Accuracy
Max. ±5 µA
Pulse/frequency output
o.r. = of reading
[in]
³⁄₈
½
1
2
3
1:10
1:20
[kg/h]
[kg/h]
200
100
650
325
1 800
900
4 500
2 250
7 000
3 500
18 000
9 000
1:10
1:20
[lb/min]
[lb/min]
7.350
3.675
23.89
11.95
66.15
33.08
165.4
82.70
257.3
128.7
661.5
330.8
Proline Promass E 100
Zero point stability
[kg/h]
[lb/min]
0.20
0.007
0.65
0.024
1.80
0.066
4.50
0.165
7.0
0.257
18.0
0.6615
1:50
1:100
[kg/h]
[kg/h]
40
20
130
65
360
180
900
450
1 400
700
3 600
1 800
1:50
1:100
[lb/min]
[lb/min]
1.470
0.735
4.778
2.389
13.23
6.615
33.08
16.54
51.46
25.73
132.3
66.15
Endress+Hauser
1:500
[kg/h]
4
13
36
90
140
360
1:500
[lb/min]
0.147
0.478
1.323
3.308
5.146
13.23

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents