Proline Promass E 100
Mounting location
Endress+Hauser
Заказывайте на сайте: www.tinnova.ru/endress_hauser/ || Эл. почта: info@tinnova.ru
Calculation of the maximum repeatability as a function of the flow rate
Flow rate
⋅
½
ZeroPoint
⋅
100
³
BaseRepeat
⋅
½
ZeroPoint
⋅
<
100
BaseRepeat
Example for maximum measured error
E [%]
2.5
2.0
1.5
1.0
0.5
0
0
5
10
15
20
25
E
Maximum measured error in % o.r. (example)
Q
Flow rate in % of maximum full scale value
Installation
No special measures such as supports etc. are necessary. External forces are absorbed by the
construction of the device.
To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid
the following mounting locations in the pipe:
• Highest point of a pipeline.
• Directly upstream of a free pipe outlet in a down pipe.
Installation in down pipes
However, the following installation suggestion allows for installation in an open vertical pipeline.
Pipe restrictions or the use of an orifice with a smaller cross-section than the nominal diameter
prevent the sensor running empty while measurement is in progress.
Maximum repeatability in % o.r.
A0021335
ZeroPoint
⋅
⋅
±
½
100
MeasValue
A0021336
30
35
40
45
50
55
60
65
70
75
80
85
90
95 100
A0021340
A0021337
Q [%]
A0030289
A0028772
45