Modbus Commands; Modbus Addressing Model; Modbus Communication - Siemens SITRANS F Coriolis FCT030 Function Manual

Flowmeters modbus from firmware 4.0
Table of Contents

Advertisement

MODBUS commands

4.1

Modbus addressing model

The device allows read/write access to one holding register block. All devices are mapped to
this Modbus address space.
4.2

Modbus communication

Table 4-1
General Modbus settings
Modbus
Data type /
register
Size in
bytes
8291
Unsigned /
2
8005
Unsigned /
2
8297
Unsigned /
2
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB
Parameter
Description
Restart com‐
Restarts the communication using
munication
configured slave address, baud rate
and parity/framing.
Slave address
DIP switch setting on the transmitter
(HW)
cassette. Address is used if DIP
switch is set to a value > 0.
Slave address
Software address of Modbus inter‐
(SW)
face. Address is used if switch is set
to 0.
Default val‐
Value range /
ue
Setting options
[units]
(units
register)
-
● 0: Cancel
● 1: Restart
-
-
1
1 - 147
4
Access
level
Write only
Read only
Read /
write
33