MODBUS commands
4.10 Modbus holding registers tables
Modbus
Data type /
register
Size in bytes
8929
Float / 4
10036
Float / 4
10038
Float / 4
8908
Float / 4
8939
Unsigned / 2
8940
Float / 4
7149
Float / 4
88
Parameter
Description
Lower range
Lower standard volume
value standard
flow value that is map‐
volume flow
ped to the lower nomi‐
nal range boundary se‐
lected by Frequency
value low (8937). Appli‐
cable if Process value
(8904) = 27.
Upper range
Upper current value
value current
that is mapped to the
upper nominal range
boundary selected by
Frequency value high
(8935). Applicable if
Process value (8904) =
25 or 26.
Lower range
Lower current value
value current
that is mapped to the
lower nominal range
boundary selected by
Frequency value low
(8937). Applicable if
Process value (8904) =
25 or 26.
Damping value
Time constant of damp‐
ing filter for frequency
output signal.
Fail-safe behav‐
Frequency output reac‐
ior
tion in case of a fault.
Fail-safe value
Output value in case of
a fault and when Fail-
safe behavior (8939) is
configured to 4.
Forced value
Forced value when
forcing is active.
Default value
Value range /
[units]
Setting options
(units register)
0.0
-
[Sm
/h]
3
(7964)
20.0
-
[mA]
4.0
-
[mA]
0.0
0.0 - 100.0
[s]
2
● 0: Frequency value
low
● 1: Frequency value
high
● 2: Last valid value
● 3: Disabled
● 4: Custom value
0.0
0.0 - 12500.0
[Hz]
1.0
0.0 - 12500.0
[Hz]
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB
Access
level
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write