Siemens SITRANS F Coriolis FCT030 Function Manual page 94

Flowmeters modbus from firmware 4.0
Table of Contents

Advertisement

MODBUS commands
4.10 Modbus holding registers tables
Modbus
Data type /
register
Size in bytes
9110
Float / 4
9112
Float / 4
9170
Float / 4
9172
Float / 4
9174
Float / 4
9176
Float / 4
94
Parameter
Description
Upper range
Upper mass flow value
value mass flow
that is mapped to the
upper nominal range
boundary selected by
Current mode (9106).
Applicable if Process
value (9102) = 0.
Lower range
Lower mass flow value
value mass flow
that is mapped to the
lower nominal range
boundary selected by
Current mode (9106).
Applicable if Process
value (9102) = 0.
Upper range
Upper process density
value density
value that is mapped to
the upper nominal
range boundary selec‐
ted by Current mode
(9106). Applicable if
Process value (9102) =
2.
Lower range
Lower process density
value density
value that is mapped to
the lower nominal
range boundary selec‐
ted by Current mode
(9106). Applicable if
Process value (9102) =
2.
Upper range
Upper medium temper‐
value tempera‐
ature value that is map‐
ture
ped to the upper nomi‐
nal range boundary se‐
lected by Current mode
(9106). Applicable if
Process value (9102) =
3.
Lower range
Lower medium temper‐
value tempera‐
ature value that is map‐
ture
ped to the lower nomi‐
nal range boundary se‐
lected by Current mode
(9106). Applicable if
Process value (9102) =
3.
Default value
Value range /
[units]
Setting options
(units register)
0.2 • max.
-
mass flow
[kg/s]
(7400)
0.0
-
[kg/s]
(7400)
Max. density
-
[kg/m
]
3
(7600)
0.08
-
[kg/m
]
3
(7600)
Max. medium
-
temperature
[°C]
(7700)
Min. medium
-
temperature
[°C]
(7700)
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB
Access
level
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write

Advertisement

Table of Contents
loading

Table of Contents