Siemens SITRANS F Coriolis FCT030 Function Manual page 34

Flowmeters modbus from firmware 4.0
Table of Contents

Advertisement

MODBUS commands
4.2 Modbus communication
Modbus
Data type /
register
Size in
bytes
8298
Unsigned /
2
8299
Unsigned /
2
34
Parameter
Description
Baud rate
Baud rate of Modbus interface.
Parity and
Parity and framing of the Modbus
framing
communication interface.
Default val‐
Value range /
ue
Setting options
[units]
(units
register)
19200 Bit/s ● 0: 9600 Bit/s
● 1: 19200 Bit/
s
● 2: 115200
Bit/s
● 3: Reserved
● 4: 38400 Bit/
s
● 5: 57600 Bit/
s
● 6: 76800 Bit/
s
● 7: 1200 Bit/s
● 8: 2400 Bit/s
● 9: 4800 Bit/s
Even pari‐
● 0: Even
ty, 1 stop
parity, 1 stop
● 1: Odd
parity, 1 stop
● 2: No parity,
2 stops
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB
Access
level
Read /
write
Read /
write

Advertisement

Table of Contents
loading