Communication

2.4 Connecting

Figure 2-6
2.4
Connecting
Note
End Of Line (EOL) termination
The FCT030 EOL termination DIP switch is default set to EOL active. To change termination
setting see Setting the EOL termination DIP switches changed.
2.4.1

Connecting the Modbus (CH1)

1. Remove cap and ferrule from cable gland and slide onto cable.
2. Push cable through open gland and cable path.
3. Restore ferrule and tighten cap to lightly hold cable in place.
4. Signal cable screen is folded back over outer sheath and grounded beneath cable clamp.
14
Maximum cable lengths in multidrop configuration
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB