Quality Codes For Process Values - Siemens SITRANS F Coriolis FCT030 Function Manual

Flowmeters modbus from firmware 4.0
Table of Contents

Advertisement

MODBUS commands
4.10 Modbus holding registers tables
Manual zero point adjustment
In case an automatic zero point adjustment cannot be performed, it is possible to do a manual
zero point adjustment by entering the zero point offset value.
1. Select Modbus address 2132 Zero Point Adjustment and set the value to 1 = Manual Zero
Point Adjustment.
2. Select Modbus address 2133 Manual Zero Point Offset and enter the desired offset value.
4.10.5.5

Quality codes for process values

Table 4-15
Quality codes for process value
Modbus
Data type /
register
Size in bytes
7409
Unsigned / 2
7509
Unsigned / 2
64
Parameter
Description
Mass flow sta‐
Quality sta‐
tus
tus code of
the process
value Mass
flow
Volume flow
Quality sta‐
status
tus code of
the process
value Vol‐
ume flow
Default value
Value range /
[units]
Setting options
(units register)
● Limit status bit 4 & 5:
● 11: Constant
● 01: Low limited
● 10: High limited
● 00: Not limited
● Process data status bit 6 & 7:
● 11: Good
● 01: Poor Accuracy
● 10: Manual / Fixed
● 00: Bad
● Limit status bit 4 & 5:
● 11: Constant
● 01: Low limited
● 10: High limited
● 00: Not limited
● Process data status bit 6 & 7:
● 11: Good
● 01: Poor Accuracy
● 10: Manual / Fixed
● 00: Bad
Function Manual, 06/2018, A5E39931813-AB
Access
level
Read
only
Read
only
FCT030 Modbus (From firmware 4.0)