The device provides means to remap 20 Modbus registers.
Table 4-3
Modbus register mapping
Modbus
Data type /
register
Size in
bytes
10448
Unsigned /
4
10450
Unsigned /
2
10451
Unsigned /
2
...
10488
Unsigned /
2
10489
Unsigned /
2
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB
Parameter
Description
Enable mapping
Activation/deactivation of
the register mapping. A set
bit means that the mapping
pair is activated, a cleared bit
that the mapping pair is de‐
activated.
Bit 0:
Requested register 1 / Tar‐
get register 1
...
Bit 19:
Requested register 20 / Tar‐
get register 20
Register 1 source
Modbus register that ap‐
pears within Modbus re‐
quest is redirected to the pa‐
rameter specified by Target
register 1
Register 1 target
Register of an existing prod‐
uct parameter to which a
Modbus request is redirected
Register 20 source Modbus register that ap‐
pears within Modbus re‐
quest is redirected to the pa‐
rameter specified by Target
register 20
Register 20 target
Register of an existing prod‐
uct parameter to which a
Modbus request is redirected
MODBUS commands
4.4 Modbus register mapping
Default value
Value range /
[units]
Setting op‐
(units register)
tions
0
0 - 1048575
65535
0 - 65535
65535
0 - 65535
65535
0 - 65535
65535
0 - 65535
Access
level
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
37