Modbus
Data type /
register
Size in bytes
8993
Unsigned / 2
8992
Usigned / 2
8994
Unsigned / 2
9276
Unsigned / 2
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB
Parameter
Description
Volume units
Units for
CH2
volume
quantities
related to
CH2 if in
pulse
mode.
Mass units CH2 Units for
mass quan‐
tities rela‐
ted to CH2
if in pulse
mode.
Standard vol‐
Units for
ume units CH2
standard
volume
quantities
related to
CH2.
Pulse width
Units of
units CH3
pulse dura‐
tion of
CHh3 in
pulse
mode.
4.10 Modbus holding registers tables
Default value
Value range /
[units]
Setting options
(units register)
41
● 40: US gallons
● 41: l (liters)
● 42: i.gal (Imperial gallons)
● 43: m³ (cubic meters)
● 46: BBL42 (1 barrel = 42 US
gallons)
● 110: bush (bushels)
● 111: yd³ (cubic yards)
● 112: ft³ (cubic feet)
● 113: in³ (cubic inches)
● 124: BBL31.5 (1 barrel = 31.5 US
gallons)
● 170: BBL31 (1 barrel = 31 US
gallons)
● 236: hl (hectoliters)
● 253: custom units (see Custom
volume units string (8452) / factor
(8450))
61
● 60: g (grams)
● 61: kg (kilograms)
● 62: t (1 t = 1000 kg)
● 63: lb (pounds)
● 64: STon (1 STon = 2000 lb)
● 65: T (1 T = 2240 lb)
● 125: oz (ounces)
● 253: custom units (see Custom
mass units string (8476) / factor
(8474))
171
● 166: Nm³ (normal cubic meters)
● 167: Nl (normal liters)
● 168: Sft³ (standard cubic feet)
● 171: Sl (standard liters)
● 172: Sm³ (standard cubic meters)
51
● 51: s (seconds)
MODBUS commands
Access
level
Read /
write
Read /
write
Read /
write
Read /
write
53