Safety and Regulatory Information

Fuses

WA RNING
WA RNING
30
The 8355 SCD and 8255 NCD require fuses for proper operation.
These must only be accessed by Agilent trained service
personnel.
For continued protection against fire hazard, replace fuses only
with fuses of the same type and rating.
Electrical shock hazard. Disconnect the instrument from the
mains supply before replacing fuses.
Table 3
AC board fuses
Fuse
Line voltage
designation
F1/F2
All
F6/F5
All
F4/F3
All
Fuse rating and type
15 A, 250 VAC, IEC type F (non-time
delay), ceramic body
0.75 A, 250 VAC, IEC type F (non-time
delay), glass body
10 A, 250 VAC, IEC type F (non-time
delay), glass body
SCD and NCD User Manual