The current software version of the PROFIBUS card and the boot software version are
indicated in the bottom lines of the display.
6.6.19

Start-up state (function 91)

Figure 6-64
This function can be used to set the device to different start-up states, such as 'Standby'
state. This setting serves to reduce usage of the pump without interrupting the process. The
currently active operating state is marked with "■", and the others with "□".
Standby
The pump is switched off with this setting. When this state is exited, the pump is started and
the device goes to 'Measuring' state.
Pause
With this setting, you reduce the gas consumption by switching off the pump. This state is
exited via the warm-up phase.
Pause without heating
With this setting, you reduce the gas consumption and power. In addition to the pump, the
heating is also switched off. This state is exited via the warm-up phase.
Note
Change of operating state
The operating state can be changed manually using function 91 or automatically as a result
of faults/maintenance requests. In section State change due to fault / maintenance request
(Page 48) of table "Change of state due to fault/maintenance request" you will find possible
state changes.