Technical specifications

9.1

General technical data of the FIDAMAT 6 gas analyzers

Table 9- 1
Technical data – Design, housing
Design, enclosure
Degree of protection
Dimensions
Weight
Table 9- 2
Technical data – Electric characteristics
Electrical characteristics
Auxiliary power supply
Power consumption
EMC interference immunity (electromagnetic
compatibility)
Electrical safety
Fuse values
Table 9- 3
Technical data – Gas consumption data
Gas consumption data
Combustion air
Hydrogen
Sample gas
Zero gas/ calibration gas
FIDAMAT 6
Operating Instructions, 03/2018, A5E00222135-03
IP20 according to EN 60529
IP40 after installation in analyzer cabinet (FIDAMAT Analyzer System)
See section "Dimension drawings and circuit diagrams"
approx. 23 kg (51 lbs.)
100 to 120 V AC (nominal range of use 90 V to 132 V), 48 to 63 Hz or
200 to 240 V AC (nominal range of use 180 V to 264 V), 48 to 63 Hz
Approx. 150 VA during operation
Approx. 350 VA in the warm-up phase
In accordance with standard requirements of NAMUR NE21 (08/98)
In accordance with EN 61010-1,
Overvoltage category II
100 to 120 V: 4.0 T/250
200 to 240 V: 2.5 T/250
Approx. 350 ml/min
Approx. 12 ml/min
Approx. 1 l/min
Approx. 1.5 l/min
9
149