6.3 Calibration
Total range calibration ("Measuring" operating state required)
A total range calibration is only possible if the
activated under function 23.
With a total range calibration, you calibrate all measuring ranges together. Specify the
"leading" measuring range under function 22. We recommend using the largest measuring
range for this.
The setpoint and current actual value of the "leading" measuring range appear in the display.
When the actual value has stabilized, you can initiate the calibration by pressing the fourth
softkey. The actual value is now made to agree with the setpoint.
If you make a mistake during the calibration (e.g. wrong calibration gas), reload the original
calibration by pressing the "CANCEL" softkey.
See also
Setting the setpoints (function 22) (Page 68)
Total / single calibration (function 23) (Page 69)
Define measuring ranges (function 41) (Page 80)
6.3.4

Setting the setpoints (function 22)

Figure 6-6
The function screen shows the setpoint input with total calibration. The fourth measuring
range is chosen here as the leading measuring range.
With single calibration, there is no choice for the leading measuring range.
68
Setting the setpoints
"Total range calibration"
Operating Instructions, 03/2018, A5E00222135-03