Diagnostic Values (Function 2); Logbook (Function 3) - Siemens FIDAMAT 6 Operating Instructions Manual

Gas analyzer for determination of total hydrocarbon content
Table of Contents

Advertisement

Functions
6.2 Analyzer status
Adapter board:
● Firmware No.:
Order No. of the software in the EPROM for the adapter board
● Software status:
Version of the adapter board software
● Boot software status:
Version of the adapter board boot software
6.2.3

Diagnostic values (function 2)

The most important diagnostic values may be called using function 2. They may allow
conclusions to be drawn for evaluation of errors or setting work.
6.2.4

Logbook (function 3)

In the logbook, all errors which led to a maintenance request (W) or a fault (S) are listed.
The limit alarm (LIM) and functional check (CTRL) are also registered. However, these do
not trigger a maintenance request or fault message.
The logbook contains a max. of eight pages, with four messages per page. It works
according to the rotating buffer principle, i.e. when all eight pages (all 32 locations) are
occupied, the oldest message is overwritten.
You can delete or block logbook entries (function 60), but you can also switch them off
individually (function 87).
Note
If an error occurs whose error message is switched off with function 87, there is no reaction
at any configured interface. This applies both to the ELAN interface and to the analog output
and the relay output.
In case of servicing, function 110 (diagnostics service) is to be observed as well. However,
this function is only accessible to service engineers.
See also
Logbook settings (function 60) (Page 93)
Error on/off (function 87) (Page 119)
List of maintenance requests (Page 131)
List of faults (Page 133)
64
Operating Instructions, 03/2018, A5E00222135-03
FIDAMAT 6