Settings for AUTOCAL check
Figure 6-15
In the
checked by Autocal check. Using functions 71 and 72, you can also select the relay output or
binary input for "Autocal check".
When a calibration limit is exceeded, the maintenance request W10 is set, and also, if
configured, the relay "Acal Chk Dif".
Both will be reset again after an error-free Autocal check. W10 is retained in the logbook.
"Start only Autocal check cyclically":
If this function is activated, the device executes an Autocal check when Autocal is started
cyclically. In other words, only the adherence to the calibration tolerances is checked, and no
calibration is performed.
Sequence:
1. Start the Autocal check:
– Using the
– Via binary input
– Via cycle
2. The device executes a sequence as configured in the
subfunction.
See also
Relay outputs (function 71) (Page 96)
Binary inputs (function 72) (Page 98)
FIDAMAT 6
Operating Instructions, 03/2018, A5E00222135-03
Autocal check
"Autocal check"
subfunction, enter the desired calibration tolerances that are to be
"Trigger autocal once"
"Acal/-check sequence"
softkey in the
"Autocal/-check sequence"
Functions
6.3 Calibration
subfunction