Linear Temperature Compensation (Function 86) - Siemens FIDAMAT 6 Operating Instructions Manual

Gas analyzer for determination of total hydrocarbon content
Table of Contents

Advertisement

Functions
6.6 Configuration
6.6.16

Linear temperature compensation (function 86)

Figure 6-61
The FIDAMAT 6 is temperature compensated both at the zero point as well as for the span.
If an additional temperature error occurs during operation, compensate it with this function.
"Aftercompensation of the zero point":
Assuming an average physics temperature, define two different correction values for ranges
of increased temperature and of decreased temperature.
"Aftercompensation of the span":
The procedure is the same as for the zero point, but the change refers to the measured
value.
Note
If the zero point deviates negatively when the temperature changes, it carried a positive sign,
and vice-versa. The same applies to the measured value.
118
Compensating temperature
Operating Instructions, 03/2018, A5E00222135-03
FIDAMAT 6