6.6 Configuration
6.6.18

PROFIBUS configuration (function 90)

This function is used to set the PROFIBUS parameters. This function can only be called if
the device contains additional PROFIBUS electronics (optional board).
Figure 6-63
Specifically, you can set the following specific PROFIBUS parameters:
● Address (PROFIBUS station address): The address can be set from 0 to 126.
● Under TAG you can display an identifier assigned to the device in the PROFIBUS
network (or the first 16 characters thereof).
● You set the configuration behavior of the device with the
Parameters 0, 1 and 3 can be selected and have the following meaning:
Parame-
ters
0
1
3
Note
In order to work with the provided GSD and EDD, the value 1 must be entered for 'Ident
number'.
● Relay on PB. This can be used to enable the 8 relays of the additional electronics (option
board) so that they can be remotely controlled via PROFIBUS. For activating, however,
none of these relays are permitted to be assigned a function.
Note
For remote control via PROFIBUS, the software version of the additional PROFIBUS
electronics (option board) must be 2.0.0 or higher.
120
Configuring PROFIBUS
Meaning
Only the PROFILE Ident number is positively acknowledged
Only the device-specific Ident number is positively acknowledged
Only the PROFILE Ident number for multivariable devices (complex analyzers) is posi-
tively acknowledged.
"Ident number"
Operating Instructions, 03/2018, A5E00222135-03
(softkey 2).