Function
Autocal check
Measurement
protection
Standby
See also
Binary inputs (function 72) (Page 98)
FIDAMAT 6
Operating Instructions, 03/2018, A5E00222135-03
Necessary control volt-
age
0 V
24 V
24 V
(min. 1 s)
N, X
N, X
The six
corrected
binary
inputs
only re-
quire one
pulse for
starting.
N, X
Remark/effect
Note:
Effective control only possible in "Measuring" operating
state!
Start Autocal check (function 24)
It is possible to define a binary input "Measurement pro-
tection" that has the following effect:
If the device is in "Measuring" operating state (function
control deactivated), it remains in this state, i.e.:
The device can no longer be decoded.
•
The device can no longer be set to "Remote".
•
The message "Measurement protection activated" ap-
pears in the status line of the measured value display.
Sets the "Standby" start-up state (Function 91). This al-
lows the sample gas pump to be switched off. Only valid
for FIDAMAT 6-E (with pump) firmware version V1.3.5 or
higher.
Functions
6.6 Configuration
101