Siemens FIDAMAT 6 Operating Instructions Manual page 103

Gas analyzer for determination of total hydrocarbon content
Table of Contents

Advertisement

Function
Autocal check
Measurement
protection
Standby
See also
Binary inputs (function 72) (Page 98)
FIDAMAT 6
Operating Instructions, 03/2018, A5E00222135-03
Necessary control volt-
age
0 V
24 V
24 V
(min. 1 s)
N, X
N, X
The six
corrected
binary
inputs
only re-
quire one
pulse for
starting.
N, X
Remark/effect
Note:
Effective control only possible in "Measuring" operating
state!
Start Autocal check (function 24)
It is possible to define a binary input "Measurement pro-
tection" that has the following effect:
If the device is in "Measuring" operating state (function
control deactivated), it remains in this state, i.e.:
The device can no longer be decoded.
The device can no longer be set to "Remote".
The message "Measurement protection activated" ap-
pears in the status line of the measured value display.
Sets the "Standby" start-up state (Function 91). This al-
lows the sample gas pump to be switched off. Only valid
for FIDAMAT 6-E (with pump) firmware version V1.3.5 or
higher.
Functions
6.6 Configuration
101