Replacement of the Developer Thermistor

[1]
Disconnect the main power.
[2]
Drain the DEVELOPER TANK or clamp the 2 TUBES to the THERMOWELL. See the illustration on
page 2-30.
[3]
Disconnect the DEVELOPER THERMISTOR at TB4-13 and TB4-14.
When you remove the DEVELOPER THERMISTOR from the THERMOWELL, a small amount of
liquid may spill. Clean any spilled solution.
[4]
Remove the DEVELOPER THERMISTOR from the THERMOWELL.
[5]
Apply SEALANT TL-3230 to the threads of the new DEVELOPER THERMISTOR.
Do not install the new DEVELOPER THERMISTOR lower than shown in the illustration.
[6]
Install the new DEVELOPER THERMISTOR 4 mm (5/32 inch) from the DEVELOPER THERMOWELL.
[7]
Connect TB4-13 and TB4-14 to the new DEVELOPER THERMISTOR.
[8]
Unclamp the 2 TUBES or fill the DEVELOPER TANK.
981777
DEVELOPER
THERMISTOR
4 mm
(5/32 in.)

Replacement of the Developer Thermistor

THERMOWELL
H112_0180BCA
H112_0180BA
2-31