Replacement Of The O-Ring - Kodak M35 Service Manual

Processor
Hide thumbs Also See for M35:
Table of Contents

Advertisement

Replacement of the O-Ring

Dangerous voltage.
[1]
Disconnect the main power.
[2]
Remove the TOP COVER and 2 SIDE
PANELS.
[3]
Install PINCH CLAMPS on the tubing of the
RECIRCULATION PUMP to prevent leakage.
[4]
Remove the:
• 4 WING NUTS
• OUTER PUMP HOUSING
• O-RING
[5]
Check that the O-RING is correctly seated.
O-RING
OUTER
WING
PUMP
NUT
HOUSING
(4)
981777
[6]
Install the OUTER PUMP HOUSING and the
4 WING NUTS.
[7]
Remove the PINCH CLAMPS.
[8]
Connect the main power.
Never operate the RECIRCULATION
PUMPS if the TANKS are empty.
[9]
Actuate the RECIRCULATION PUMPS.
[10]
Check for agitation on the surface of the
solutions.
[11]
Check for leakage.
IMPELLER
WASHER
SCREW
SHAFT
H112_0084DCA
H112_0084DA