Adjustment Of The Replenishment Flow Rates In M35A Processors With Serial No. 8940 And Below Or In M35 Processors With Serial No. 3760 And Below - Kodak M35 Service Manual

Processor
Hide thumbs Also See for M35:
Table of Contents

Advertisement

Adjustment of the Replenishment Flow Rates
in M35A Processors with Serial No. 8940 and Below or
in M35 Processors with Serial No. 3760 and Below
[1]
Remove the:
• 2 CROSSOVER ASSEMBLIES
• developer and fixer RACK ASSEMBLIES
[2]
Insert the J-TUBE into the developer TANK.
[3]
Energize the processor.
[4]
Manually actuate the DETECTOR SWITCH by
lifting the top DETECTOR ROLLER of the
DETECTOR CROSSOVER ASSEMBLY.
[5]
Allow the developer to fill the J-TUBE.
[6]
Release the DETECTOR ROLLER.
[7]
Hold a GRADUATE under the J-TUBE.
H112_0142BCA
H112_0142BA
2-40
Adjusting the Replenishment Flow Rates
in M35A Processors with Serial No. 8940 and Below or
in M35 Processors with Serial No. 3760 and Below
[8]
To adjust the developer flow rate, lift the top
DETECTOR ROLLER to operate the
REPLENISHMENT PUMP for 28 or
34 seconds. See the table on page 2-41.
The REPLENISHMENT PUMP
operates for 34 seconds for each 35 x
43 cm (14 x 17 inch) sheet of film.
[9]
Check that the amount of developer in the
GRADUATE is the same as in the table.
[10]
Check the developer flow rate again by doing
steps 4 - 9.
[11]
Clean the J-TUBE before inserting it into the
fixer TANK.
[12]
To adjust the fixer flow rate:
(a)
Insert the J-TUBE into the fixer TANK.
(b)
Do steps 4 - 9.
GRADUATE
NOTE
TANK
J-TUBE
981777