Options for Temperature Measurement Unit are shown in Table 5-10 .
Table 5-10
Options for Temperature Measurement Unit
Description
Degrees Celsius
Degrees Fahrenheit
Degrees Rankine
Kelvin
5.7.2
Configure Temperature Damping
Display
Not available
ProLink II
ProLink→Configuration→Temperature→Temp Damping
Field Communicator
Configure→Manual Setup→Measurements→Temperature→Temp Damping
Damping is used to smooth out small, rapid fluctuations in process measurement. The Damping Value
specifies the time period (in seconds) over which the transmitter will spread changes in the reported
process variable. At the end of the interval, the reported process variable will reflect 63% of the change
in the actual measured value.
Tips
• A high damping value makes the process variable appear smoother because the reported value
must change slowly.
• A low damping value makes the process variable appear more erratic because the reported
value changes more quickly.
Procedure
Enter the desired value for Temperature Damping.
The default value is 4.8 seconds. The range is 0.0 seconds to 38.4 seconds.
When you enter a value for Temperature Damping, the transmitter automatically rounds it down to the
nearest valid value. Valid values for Temperature Damping are 0, 0.6, 1.2, 2.4, 4.8, ... 38.4.
Effect of Temperature Damping
Temperature Damping affects the response speed for temperature compensation. Temperature
compensation adjusts process measurement to compensate for the effect of temperature on sensor
tube stiffness.
Configuration and Use Manual
Display
°C
°F
°R
°K
Label
ProLink II
degC
degF
degR
degK
Configure process measurement
Field Communicator
degC
degF
degR
Kelvin
89