Restrictions
• For some alarms, Status Alarm Severity is not configurable.
• For some alarms, Status Alarm Severity can be set only to two of the three options.
Procedure
1.
Select a status alarm.
2.
Set Status Alarm Severity as desired.
Option
Fault
Informational
Ignore
Status alarms and options for Status Alarm Severity
Each status alarm has a default Status Alarm Severity. Some status alarms can be configured for other
severity levels.
Configuration and Use Manual
Configure device options and preferences
Description
Actions when alarm is detected:
• Alarm posted to Alert List
• Outputs go to configured fault action (after
Fault Timeout has expired, if applicable)
• Digital communications go to configured
fault action (after Fault Timeout has expired,
if applicable)
• Status LED changes to red or yellow
(depending on alarm severity)
Actions when alarm clears:
• Outputs return to normal behavior
• Digital communications returns to normal
behavior
• Status LED changes behavior (returns to
green, may or may not flash)
Actions when alarm is detected:
• Alarm posted to Alert List
• Status LED changes to red or yellow
(depending on alarm severity)
Actions when alarm clears:
• Status LED changes behavior (returns to
green, may or may not flash)
No action
109