Configure The Petroleum Measurement Application - Emerson Micro Motion 9739 MVD Configuration And Use Manual

Hide thumbs Also See for Micro Motion 9739 MVD:
Table of Contents

Advertisement

Calibration Pressure is the pressure at which your sensor was calibrated, and therefore defines the
pressure at which there will be no pressure effect. Calibration Pressure for your sensor is provided on
the sensor calibration sheet. If the data is unavailable, enter 20 PSI.
5.
Decide how pressure data will be provided to the transmitter, and perform the required setup.
• If you will poll an external pressure device, set up polling for pressure.
• If you will use a static pressure value, set Pressure Units to the units you are using, enter
External Pressure, and ensure that polling for pressure is disabled.
• If you will use digital communications or a direct analog current to write pressure data to the
transmitter, set Pressure Units to the units to be used and ensure that polling for pressure
is disabled. Then ensure that the appropriate value is written to transmitter memory at
appropriate intervals.
• If you will use an external measurement device, configure the mA input to External Pressure.
You must also enable External Pressure Compensation and set the Pressure Units to the units set at
the external measurement device.
5.9

Configure the petroleum measurement application

Display
Not available
ProLink II
ProLink→Configuration→API Setup
Field Communicator
Configure→Manual Setup→Measurements→Petroleum Measurement
The petroleum measurement parameters control the values that will be used in the transmitter's
petroleum measurement application.
The petroleum measurement parameters include:
• API Table Type
• Thermal Expansion Coefficient (TEC) (if required by API Table Type)
• Reference Temperature (if required by API Table Type)
Restriction
The petroleum measurement parameters are available only if the petroleum measurement application has
been purchased and is enabled on your transmitter.
Procedure
1.
Select API Table Type.
2.
If you set API Table Type to 53A, 53B, 53D, or 54C, set Reference Temperature to the appropriate value for
your application. Enter the value in °C.
3.
If you set API Table Type to 6C, 24C, or 54C, set Thermal Expansion Coefficient to the appropriate value
for your application.
4.
(Optional) Set the temperature unit configured on the transmitter to the temperature unit used by
your API reference table.
Configuration and Use Manual
Configure process measurement
91