Emerson Micro Motion 9739 MVD Configuration And Use Manual page 72

Hide thumbs Also See for Micro Motion 9739 MVD:
Table of Contents

Advertisement

Configure process measurement
Table 5-1
Options for Mass Flow Measurement Unit
Unit description
Short tons (2000 pounds)
per hour
Short tons (2000 pounds)
per day
Long tons (2240 pounds)
per hour
Long tons (2240 pounds)
per day
Special unit
Define a special measurement unit for mass flow
Display
Not available
ProLink II
ProLink→Configuration→Special Units
Field Communicator
Configure→Manual Setup→Measurements→Special Units→Mass Special Units
A special measurement unit allows you to report process data, totalizer data, and inventory data in a
unit that is not hard-coded in the transmitter. A special measurement unit is calculated from an existing
measurement unit using a conversion factor.
Restriction
Although you cannot define a special measurement unit using the display, you can use the display to
select an existing special measurement unit and to view process data.
Procedure
1.
Specify Base Mass Unit.
Base Mass Unit is the existing mass unit that the special unit will be based on.
2.
Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3.
Calculate Mass Flow Conversion Factor as follows:
a.
x base units = y special units
b.
Mass Flow Conversion Factor = x/y
4.
Enter Mass Flow Conversion Factor.
5.
Set Mass Flow Label to the label to be used for the mass flow unit.
6.
Set Mass Total Label to the label to be used for the mass total and mass inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the
special measurement unit at any time.
66
continued
Display
ST/H
ST/D
LT/H
LT/D
SPECL
Label
ProLink II
sTon/hr
sTon/day
lTon/hr
lTon/day
special
Micro Motion 9739 MVD Transmitters
Field Communicator
STon/h
STon/d
LTon/h
LTon/d
Spcl

Hide quick links:

Advertisement

Table of Contents
loading